Optimization and Validation of Reversed Phase Ion-Pair Liquid Chromatographic Method for Accurate Determination of Nitrate in High Saline Seawater | ||||
Egyptian Journal of Chemistry | ||||
Volume 68, Issue 4, April 2025, Page 135-139 PDF (358.19 K) | ||||
Document Type: Original Article | ||||
DOI: 10.21608/ejchem.2024.308012.10094 | ||||
![]() | ||||
Authors | ||||
Ibrahim Farag Tahoun ![]() ![]() | ||||
1National Institute of Standards, Tersa St, El-Matbah, Haram, P. O. Box: 136 Giza, Code No: 12211 Giza, Egypt | ||||
2Organic Analysis Laboratory, Chemical Metrology Division, National Institute of Standards, Giza, Egypt | ||||
Abstract | ||||
An accurate method for the determination of low concentrations of nitrate in high-saline seawater is now well optimized and validated. The method is based on using reversed-phase ion-pair liquid chromatography with diode-array detection for accurate determination of nitrate in high-saline seawater without interference from highly dominant chloride and sulfate anions. The method showed satisfactory precision (ranging from 5.8% to 9.5%), linearity, and detection limits, and its trueness was confirmed by participation in the international comparisons of CCQM K-161 with seven national metrology institutes. The key comparison and spiking showed recovery rates between 90% and 110%. The validation results confirm the applicability of the optimized method for detecting very low concentrations of nitrate in high-saline seawater. | ||||
Keywords | ||||
Nitrate; High Saline Seawater; Method validation; Ion-Pair HPLC | ||||
Statistics Article View: 1,022 PDF Download: 145 |
||||