EFFECT OF VARIATIONS OF FIELD OF VIEW (FOV) AND MILLIAMPERE (MA) ON METAL ARTIFACTS IN CONE BEAM COMPUTED TOMOGRAPHY (IN VITRO STUDY) | ||||
Advanced Dental Journal | ||||
Volume 6, Issue 4, October 2024, Page 656-665 PDF (415.14 K) | ||||
Document Type: Original Article | ||||
DOI: 10.21608/adjc.2024.243943.1419 | ||||
![]() | ||||
Authors | ||||
Roshan Adel Ahmed ![]() ![]() ![]() ![]() ![]() | ||||
1Oral and Maxillofacial Radiology Department, Faculty of Dentistry, Cairo University, Egypt. | ||||
2Oral and Maxillofacial Radiology Department, Faculty of Dentistry, Beni Suef University, Egypt. | ||||
Abstract | ||||
Objective: The aim of this study is to evaluate the effect of exposure parameters as milliampere (mA) and field of view (FOV) in CBCT on metal artifact of dental implant, Can change of milliampere (mA) and field of view (FOV) reduce the metallic artifacts in CBCT that result from metallic structures? Methodology: Four dry human mandibles with implants inserted in them, were scanned by using Planmeca ProMax® 3D Mid CBCT machine, using different values of FOV and mA, without the activation of MAR tool. Romexis® software was used for image analysis. For each implant linear measurements were measured (length & width), implants real length and width were used as gold standard. Results: In general, the change of the size of FOV had no significant effect on metal artifacts reduction (P > 0.05), Also; it was the same on mA values, there was no significant effect on metal artifacts reduction (P > 0.05). Conclusions: According to the results of this study, Size of FOV and different values of mA of CBCT have no significant effect on the metal artifacts around dental implants. | ||||
Keywords | ||||
CBCT; CBCT exposure parameters; Metal artifacts; beam hardening artifacts; Dental implant | ||||