Comparison of Intraocular Pressure Measurement by Goldmann Applanation Tonometer and Air Puff Tonometer in Silicone Filled Eyes. | ||||
Benha Medical Journal | ||||
Article 17, Volume 37, Issue 3, September 2020, Page 667-677 PDF (595.35 K) | ||||
Document Type: Original Article | ||||
DOI: 10.21608/bmfj.2020.109131 | ||||
View on SCiNiTO | ||||
Authors | ||||
Ashraf El Shayeb1; Mohamed Elmohamady1; Noha Kilany2 | ||||
1Department of Ophthalmology, Benha faculty of medicine, Benha University, Egypt. | ||||
2Department of Ophthalmology, Benha faculty of medicine, Benha University, Egypt | ||||
Abstract | ||||
Background: To compare intraocular pressure (IOP) measurements in silicone filled eyes by Goldmann applanation tonometer (GAT) and air puff tonometer (APT). Aim: The purpose of this study was to evaluate the role of APT by comparing the measurements of IOP made using it with those made using a GAT. Methods: In this prospective comparative study, 40 eyes with silicone oil tamponade were included. IOP was measured by GAT and air puff tonometer at least 1 month from vitrectomy. Results: in eyes with silicone oil, IOP was 21.35±12.34 mmHg and 19.46±5.29 mmHg using GAT and APT, respectively (p>0.05) and this difference was not statistically significant (p=0.074). In addition, no significant correlation was detected between IOP measurements using both techniques and gender, and lens status. Conclusion: It seems that GAT provided higher IOP measurements by 3.34 mmHg (95% CI = 2.15 to 4.53 mmHg) on average compared with APT but the difference between the two IOP measurements was statistically non-significant (p >0.05), in silicone filled eyes. The further assessment of available methods for IOP measurement could possibly establish the most accurate technique for IOP estimation in silicone filled eyes. | ||||
Keywords | ||||
Intraocular pressure; Goldmann applanation tonometry; Air puff tonometry; silicone oil | ||||