Optical Biometry versus Applanation Ultrasound Biometry in Axial Eye Length Measurement in Pseudophakia | ||||
The Egyptian Journal of Hospital Medicine | ||||
Article 12, Volume 67, Issue 1, April 2017, Page 377-381 PDF (754.21 K) | ||||
Document Type: Original Article | ||||
![]() | ||||
Author | ||||
Hayam Sayed Kamel | ||||
Ophthalmology Department, Faculty of Medicine (for Girls) , Al – Azhar University, Cairo, Egypt | ||||
Abstract | ||||
Purpose: to compare axial eye length (AEL) measurement in pseudophakic eyes by optical biometry and applanation ultrasound (u/s) biometry. Methods: a prospective, case controlled and comparative study was performed at Al-Zahraa University Hospital clinic from (January 2016 – December 2016). In total 45 eyes were enrolled; 24 pseudophakic eyes were attending the clinic for routine post operative follow up; and 21 eyes with clear crystalline lens as a control group. AEL was measured in both groups using both optical biometry and applanation u/s biometry. Results & Conclusions: both optical biometry andapplanation u/s biometry show no significant difference in AEL measurement in pseudophakic eyes as well as the control group. | ||||
Keywords | ||||
Pseudophakia; Ultrasound biometry | ||||