Development of EPR dosimeters based on SiO2 for possible use in radiation processing applications | ||||
Egyptian Journal of Solids | ||||
Article 1, Volume 34, Issue 2, 2011, Page 1-18 PDF (480.92 K) | ||||
DOI: 10.21608/ejs.2011.148867 | ||||
View on SCiNiTO | ||||
Abstract | ||||
Quartz rods (310 mm) for routine use in high-dose radiation applications have been prepared by a simple technique in the laboratory where Quartz powder of two sources q1 and q2 each was mixed with molten mixture of paraffin wax and ethylene vinyl acetate copolymer (EVA). The binding mixture of EVA/Paraffin does not exhibit interference or noise in the EPR signal before or after the irradiation. The rods show good mechanical properties for safe and multi-use handling. The rods can be used with good precision in the dose range from 0.1 to 15 kGy for q1 and from 0.1 to 25 KGy for q2. The overall uncertainty for calibration using EVAPA rod program was found to be 4.55 % for q1 and 5.95 % for q2.The dose response, influence of humidity and temperature during irradiation, energy dependence as well as post-irradiation storage at different conditions are discussed. | ||||
Keywords | ||||
Quartz, Ethylene Vinyl Acetate, Radiation Dosimetry; EPR | ||||