APPLICATION OF HPLC FOR THE ANALYSIS OF MALATHION AND S-MUSTARD MIXTURE | ||||
The International Conference on Chemical and Environmental Engineering | ||||
Article 32, Volume 4, 4th International Conference On Chemical & Environmental Engineering, May 2008, Page 515-522 PDF (313.71 K) | ||||
Document Type: Original Article | ||||
DOI: 10.21608/iccee.2008.38445 | ||||
![]() | ||||
Authors | ||||
KASSEM M1; FAYED M.1; Soltan A.1; El Ahmad L. A.2 | ||||
1Egyptian Army Forces. | ||||
2Syrian Army Forces. | ||||
Abstract | ||||
ABSTRACT A method was developed for the chromatographic separation and quantification for a mixture of Malathion (as organophosphorus nerve agent), and S-mustard (as blister agent). The method used a high performance liquid chromatography (HPLC) with reversed phase C-18 column, and UV detection at 220 nm. The compounds were separated using isocratic elution acetonitrile-water (50%) at a flow rate of 0.7 mL/min in a period of 5 min. The retention times were (1.939) min for S-mustard and (3.063) min for Malathion. The calibaration curves were linear (R2 = 0.999) in the studied concentration range (200 to 1000 μg/mL). Sensitivity of the method for quantification of Malathion was found to be higher than that of S-mustard by four times. In addition, the accuracy was found to be (~ 97%) for both. | ||||
Keywords | ||||
malathion; S-mustard; HPLC analysis | ||||
Statistics Article View: 96 PDF Download: 422 |
||||