Frictional resistance of active and passive aesthetic self-ligating brackets with different orthodontic wire sizes | ||||
Egyptian Orthodontic Journal | ||||
Article 5, Volume 52, December 2017, December 2017, Page 45-54 PDF (2.51 MB) | ||||
Document Type: Original Article | ||||
DOI: 10.21608/eos.2017.78358 | ||||
View on SCiNiTO | ||||
Authors | ||||
Yasser M Emam* ; Sherif Morcos; Mohamed A Nadim; Abbadi A El Kady | ||||
Orthodontics department, Faculty of Dentistry, Suez Canal University, Ismailia, Egypt | ||||
Abstract | ||||
The purpose of this study was toinvestigate the effect of passive and active Self-ligation of aesthetic brackets on the static and kinetic friction with different wire sizes. Two types of ceramic brackets were used; Empower Clear (active self-ligating bracket (ASLB)) and Damon Clear 2 (passive self-ligating bracket (PSLB)). They were coupled with 0.017 x 0.025-in and 0.019 x 0.025-in stainless-steel (SS) wires. A universal testing machine was used to pull the brackets along the distal end of the archwire. The test was performed at 5 mm/min crosshead speed over a distance of 8 mm to measure the static friction (SF) and the kinetic friction (KF) under dry conditions. The results showedthe PSLB’s to have significantly lower SF and KF with both wire sizes. | ||||