Comparison of Strain Development of Implant Supported Three Units Fixed Partial Dentures Fabricated from Zirconia and PEEK “In-vitro study” | ||||
Egyptian Dental Journal | ||||
Volume 70, Issue 3 - Serial Number 4, July 2024, Page 2537-2547 PDF (1.59 MB) | ||||
Document Type: Original Article | ||||
DOI: 10.21608/edj.2024.280626.2994 | ||||
![]() | ||||
Authors | ||||
Mostafa hassan![]() ![]() ![]() ![]() ![]() | ||||
1Instructor, Faculty of Oral and Dental Medicine Misr International University | ||||
2Professor, Fixed Prosthdontic Department, Faculty of Oral and Dental Medicine, Misr international University, Cairo, Egypt | ||||
3Professor, Chairperson Head of Fixed Prosthodontics, Conservative Dentistry Department, Faculty of Oral and Dental Medicine, Misr International University (MIU), Cairo, Egypt | ||||
4Assistant Professor, Fixed Prosthodontics Division, Conservative Department, Faculty of Oral and Dental Medicine, Misr International University, Cairo ,Egypt | ||||
Abstract | ||||
Statement of the Problem: selection of prosthetic materials is a critical factor affecting the long-term success of implant supported restorations. PEEK poly-ether-ether ketone-is a quite new polymer used for dental applications. However, there are limited data available concerning Stress analysis and strain development of PEEK restorations as implant superstructure. Introduction; Materials and Methods; A total of 14 frameworks (n=7) were fabricated of Zirconia and PEEK by milling technique, divided into two equal groups to be randomly seated on its corresponding titanium abutment and then subjected to 5000 thermo-cycles. Strain gauge method was selected in this study to assess strain development where strain gauges were adhesively bonded mesial and distal to each implant-abutment unit and data were recorded using a strain meter device. Results; As regards the overall strain (mean of the four channels): there was no statistically significant difference between the two groups Conclusion; PEEK frameworks revealed slightly higher strain compared to zirconia with no significant difference in the over-all strain channels. | ||||
Keywords | ||||
implant; PEEK; strain; Fixed | ||||